高效液相色谱法测定青霉素钠含量的测量不确定度评定
林小莉
0
(哈尔滨市兽药饲料监察所)
摘要:
采用高效液相色谱法测定青霉素钠的含量,评定其不确定度。建立测量不确定度的数学模型,通过对整个测定过程中所产生的各不确定度分量的分析和计算,求得当青霉素钠的含量96.6%时,扩展不确定度为2.7%。
关键词:  测量不确定度  高效液相色谱法  青霉素钠  评定
DOI:
投稿时间:2008-09-06修订日期:2008-10-17
基金项目:
Evaluation of Measurement Uncertainty in the Determination of Benzylpenicillin Sodium by High Performance Liquid Chromatography
LIN Xiaoli
(Institute of Veterinary Drug and Food Control in Harbin)
Abstract:
The uncertainty in the determination of benzylpenicillin sodium by High Performance Liquid Chromatography was evaluated. The mathematical model was established. By analyzing and quantifying the components of uncertainty, the expanded uncertainty of 2.7% with 96.6% of the content of benzylpenicillin sodium was derived.
Key words:  measurement uncertainty  High Performance Liquid Chromatography  benzylpenicillin sodium  evaluation  benzylpenicillin sodium  evaluation